ILT-U-1872
Vibration testing
DETERMINATION | METHOD |
| Vibration (Level 1)
o Minimum frequency: 10 Hz. o Frequency f1: 20 Hz Frequency f2: 1000 Hz o Maximum frequency: 2000 Hz. o RMS value: 3.35 g o ASD form (flat between f1 and f2): 0.01 g2 / Hz o Initial slope of 10 Hz: +6 dB / octave o Final slope between 1000 and 2000 Hz: -24 dB / octave o Duration: 20 minutes
| IEC 60068-2-64 |